Step 5. Verify the Triple Quadrupole LC/MS sensitivity
The Performance Specification Services (G6061A, G6062A, G6063A, G6067A,
G6068A, and G6069A) can be purchased only at the time of original Agilent
6400 Series Triple Quadrupole LC/MS System bundle purchase.
• For standard Marketing Bid Specification Services (G6061A for Positive Ion
• For the Instrument Detection Limit (IDL) Marketing Bid Specification
Figure 30
Reserpine checkout worklist
1 Review the worklist to be sure that the method and data paths are correct.
2 Run the Negative Ion Mode checkout worklist.
Agilent 6400 Series Triple Quad LC/MS System Installation Guide
Mode, G6062A for Negative Ion Mode, G6063A for both Negative and
Positive Ion Modes), the same checkout conditions as the installation
checkout methods are used.
Services (G6067A for Positive Ion Mode), same checkout conditions as the
installation checkout methods are used, but with different sample dilutions.
Check that the data file names given in the worklist are unique and have not
already been acquired.
After you run the checkout worklist in Negative mode, you will repeat this
step in Positive mode. This step includes instructions for both modes. Make
sure you use the correct instructions.
a Run the appropriate Negative Ion Mode checkout worklist.
b Load each solvent blank and chloramphenicol sample data file (or
reserpine peak for Positive Ion Mode) into the Qualitative Analysis
program.
c Generate Extracted Ion Chromatograms of the ion product (152.0 m/z for
Negative Ion Mode, 195.10 m/z for Positive Ion Mode).
Installation Completion and System Verification
Step 5. Verify the Triple Quadrupole LC/MS sensitivity
3
63