6
Reference

Positive Ion Mode Checkout Conditions

Positive Ion Mode Checkout Conditions
These conditions are used in a Positive Ion Mode checkout.
Table 11
Time
Initial
0.1
1
2
2.01
4
Table 12
Injection volume
Autosampler temperature
Needle wash flushport
Needle wash location
106
HPLC Pump
%A (water w/0.1%
formic acid)
90
90
2
2
90
Stop
HPLC Autosampler
Agilent 6400 Series Triple Quad LC/MS System Installation Guide
%B (acetonitrile)
Flow
10
0.4 mL/min
10
0.4 mL/min
98
0.4 mL/min
98
0.4 mL/min
10
0.4 mL/min
2 µL
4°C
75:25 methanol:water
Flushport, 10 seconds