2
Software Installation and Configuration
Step 5. Install the Qualitative Analysis program
4 If your system includes an Agilent 1290 Infinity LC Injector HTC/HTS, click
5 Update the firmware:
Step 5. Install the Qualitative Analysis program
• Refer to the Offline Qualitative and Quantitative Analysis Installation
You must use the Run as administrator option to install the program. Failure to do so
C A U T I O N
can result in an incomplete installation, which can cause the system to run unreliably.
Do not install any other MassHunter program until instructed to do so.
40
•
If you are asked to allow Agilent Technologies to install software on the
computer, click Yes.
•
If you get a message that indicates that the screen resolution is not set to
the required resolution, click OK.
•
If you see a Notepad window that prompts you to install other options,
close the Notepad window.
•
When the installation completion message is displayed, click Finish.
Install HTC/HTS Driver.
a Insert the Firmware installation media.
b Click Update MS Firmware in the welcome page.
Follow the instructions to update the firmware.
Guide to install the Qualitative Analysis program.
Agilent 6400 Series Triple Quad LC/MS System Installation Guide