To reconfigure the instrument
If you have more than one stack of HPLC modules from which to choose, or a
new supported LC module is installed, the Instrument Configuration registry
must be cleared for MassHunter Workstation.
1 Put the Triple Quadrupole LC/MS into standby mode.
2 Close the MassHunter Acquisition program.
3 Stop MassHunter processes.
4 Start the Instrument Configuration program (Programs > Agilent >
5 Mark the Mass Spectrometer check box, and then click the Triple
Figure 37
Agilent 6400 Series Triple Quad LC/MS System Installation Guide
See
"To stop MassHunter processes in case of unrecoverable errors"
page 85.
MassHunter Workstation > Acq Tools > Instrument Configuration).
Quadrupole LC/MS. Leave all other check boxes cleared.
Instrument Configuration
Other Setup Tasks
To reconfigure the instrument
5
on
89