13 Troubleshooting
Errors and warnings in the scan
progress window
Error message
Overexposure error occurred. Set
a lower PMT voltage.
Scanning stopped because the
door was opened.
The combination of the laser and
filter might be inappropriate.
Check the laser and filter.
A scanner error was detected.
Please restart Typhoon FLA 7000
imager and Scanner Control Soft-
ware.
Sense Key : %1H
Error Code : %2H(%3H)
96
Meaning and countermeasure
The scanned image was overexposed due to
a high PMT voltage setting.
Countermeasure: Decrease the PMT voltage
in the Reader Settings window.
The door was opened during the scan.
Countermeasure: Close the door and scan the
image again. Do not open the door until the
scanning is finished.
The selected method may be inappropriate.
Countermeasure: Select another method, or
edit the method to suit the current application.
An error occurred during scanning.
Countermeasure: Try to run the scan again.
Contact your GE Healthcare representative if
the problem persists.
Typhoon FLA 7000 User Manual 28-9607-62 AC