Step
4
5
6
Typhoon FLA 7000 User Manual 28-9607-62 AC
Action
Place the sample on the cassette. Make sure that the sample surface faces
up.
Keep the radio imaging sample away from the edge of the
Note:
storage phosphor screen. Otherwise, the recorded data
may become corrupted.
Place the storage phosphor screen in the cassette, with the exposure surface
of the storage phosphor screen facing the sample. Make sure that the notch
of the storage phosphor screen is in the front left corner of the cassette as
shown below.
Close the cover of the cassette. Press on the cover until it clicks into place.
11 About storage phosphor screens
11.5 Exposing the storage phosphor screen