Error message
A laser error was detected.
Please retry the scan, or
contact service.
Please select image folder.
Please input a file name.
Please set PMT Voltage value
to 500-1000.
Please give a name to the
condition.
Failed to retrieve image data
from the scanner. Check the
scanner connection.
The stage is not properly in-
serted. Please insert the
stage in its correct position.
Typhoon FLA 7000 User Manual 28-9607-62 AC
Meaning and countermeasure
There was an error with the laser.
Countermeasure: Try to scan again.
Countermeasure: Restart the instrument and Ty-
phoon FLA 7000 Control Software, then try to run
the scan again.
Countermeasure: Scan using a different laser.
No image folder is selected.
Countermeasure: Select an image folder by clicking
the Browse... button and navigating to a suitable
folder.
No name has been assigned to the image file.
Countermeasure: Type a name in the File Name
field.
The PMT voltage setting is outside the permitted
range.
Countermeasure: Type a PMT voltage value be-
tween 500 and 1000 in the PMT field.
No name was given to a condition before saving.
Countermeasure: Type a name in the Condition
Name field.
The image data could not be retrieved from the
scanner.
Countermeasure: Check that the scanner is properly
connected to the computer, then try to run the scan
again.
Countermeasure: Disconnect all other USB devices
from the computer and run the scan again.
Countermeasure: Restart the computer and Ty-
phoon FLA 7000 Control Software, then try to run
the scan again.
The stage is not properly inserted.
Countermeasure: Insert the stage properly.