10 Reading conditions, display parameters, and other settings
10.1 Reading conditions explained
Latitude
This setting specifies the dynamic range. The dynamic range that can be detected is
larger with L5 than with L4. If the signals of the sample are in the L4 range, the density
gradation is represented in more detail if L4 is selected.
Stage
Select the stage which will be used in the scan. The image under Sampling Area changes
to reflect the selection.
Sampling area
Displays the area or areas to be scanned as red boxes.
Mode
Sets the selection behavior of Sampling area.
Setting
Grid
Free
All
File Size
This area displays the estimated size of the image file as determined by the current
settings.
80
Change the size of the area or areas to be scanned by clicking and dragging the
edges of the red box with the mouse.
Change the position of the area or areas to be scanned by clicking and dragging the
red box with the mouse.
Description
The sampling area is increased or decreased in steps corresponding to
the grid of the stage or the positions of the titer plate plugin.
The sampling area is increased or decreased in a stepless fashion.
The sampling area is expanded to the whole readable surface of the stage
or titer plate plugin is selected.