Typhoon FLA 7000 User Manual 28-9607-62 AC
Part
Description
6
Set the scanning area.
1 Select a stage in the drop-down menu.
2 Drag the red square to the desired position of the scanning
area.
3 Drag the sides of the red square as needed to adjust the size
of the scanning area.
7
Pixel Size: set the pixel size.
Choose a small pixel size for high quality images. Note that a small
pixel size setting increases the reading time and the size of the
image file.
8
Latitude: Specify the dynamic range.
L5 provides a larger detectable range than L4.
•
L4 provides a finer density gradation, if the signals of the sample
•
areas are in the correct range.
9
Mode: Specify the mode for the reading area borders.
Select Grid to limit the borders to the grid lines.
•
Select Free for an arbitrary area.
•
Select All to use the entire Fluor stage as the reading area.
•
10
Save Condition...: click this button to save the current reading
conditions in a file, if desired. For details, refer to the User Manual.
11
Load Condition...: click this button to load previously saved reading
conditions, if desired. For details, refer to the User Manual.
12
File Size: the estimated size of the result data file is presented.
13
Reading Time: the estimated time required for the scan is present-
ed.
14
Top: return to the main window.
15
Start Scan: start the scan. The sample must be loaded before
starting a scan.
1.4 Typhoon FLA 7000 Control Software
1 Introduction
13