IEEE-488 PROGRAMMING
4.9.11
Voltage Source (V)
Purpose
Format
Parameters
Used to set the DC voltage level of the voltage source.
Description
Default
Reference
Example
Vnn.nn
nn.nn=Lmv Level (L.) to High Level (H)
NOTE Only as many significant digits as necessary need be sent (i.e. send V4 in
place of VO4.00).
The range of the voltage source is
from
-2O.OOV to 2O.OOV. When making capacitance
measurements, the square wave or staircase originates from the programmed DC
voltage level. Note that the V cormnan d does not apply the programmed voltage
level to the voltage source output connector. The output is controlled by the W
(waveform) command (see paragraph 4.9.14).
Upon power up or after a DCL or SDC command is received, the voltage source
will be programmed to OO.OOV.
The equivalent front panel controls for the V command are the DISPLAY SOURCE;
ADJUST end +I- buttons. See paragraphs 3.3.10,3.3.11 and 3.3X!, respectively for
complete information on the voltage source.
Cycle power on the Model 595 and enter the foliowing statements into the computer
toset the voltage source to l5Vz
After END LINE is pressed the second time, the present voltage source setting will
be displayed. After END LINE is pressed the third time, the voltage source value
will be programmed tom l5V.
4-2.5