IEEE488 PROGRAMMING
e.g.3 Disday (D)
Purpose
Format
Parameters
Determines the Model 595 data displayed on the front panel.
Dn
n = 0 Meter
n = 1 Voltage Source
Description
Default
Reference
n = 2 ?Iigh Limit of Voltage Source
n = 3 Low Limit of Voltage Source
n = 4 Step Voltage
n = 5 Delay Tie
The instrumeiit czmbe programmed to display capacitance/current measurements,
the voltage source value, the high or low limit settings of the voltage source, the
selected step voltage, or the delay time setting.
Upon power up or after receiving a DCL or SDC command, the instrument will be
in DO.
In general, the equivalent~front panel controls for the D command are the LIMITS,
STEP V, DELAY TIME and DISPLAY SOURCE buttons. See paragraphs 3.3.7 through
3.3.10 for front panel cofifrol of the display.
From the front panel, press the CURRENT or CAFACDANCE button. To display the
high limit of the voltage source, enter the following program statements into the
computer:
When END LINE is pressed the second time, the high limit of the voltage source
will be displayed. The last statement returns the instruments to the meter display.