Commissioning
7.3 5-minute quick start with the operator panel and the Ready-for-Use software
7.3.5

Setting calibration points

Figure 7-5
Carry out the calibration commands at the end of the quick start:
1. Carry out the "Set calibration weight 0" command. The "Calibration weight 0" defined in
step 4 is now visible in the display.
2. Place the "Calibration weight 1" defined in step 4 on the scale construction, and execute
the "Set calibration weight 1" command.
3. If an "Calibration weight 2" was selected:
Place "Calibration weight 2" defined in step 4 on the scale construction, and execute the
"Set calibration weight 2" command.
4. Calibration of the scale is now complete. Return to the start screen by clicking on the
house icon.
36
Quick start step 5a
SIWAREX WP321
Manual, 07/2014, A5E33715669A-02