To Set Up The Power Meter; To Start The User Flatness Cal - HP 83751A User Manual

Synthesized sweepers
Hide thumbs Also See for 83751A:
Table of Contents

Advertisement

Creating User Flatness Arrays
Creating a user flatness array for use in a scalar analysis measurement
Attention!
Do not proceed until the HP 8757 system interface is turned off by pressing @FiiZ]. Then by
pressing softkeys, MORE, SWEZP
1. Zero and calibrate the power meter/sensor.
2. Enter the appropriate power sensor calibration factors into the power
meter.
3. Enable the power meter/sensor cal factor array. For operating information
on the HP 437B power meter refer to its operating and service manual.
4. Connect the power sensor to the point where corrected power is desired.
(See Figure 5-2.)
1. On the sweeper, press m FLTMESS CAL
a. Select whether to calibrate over St>arSt.op (correction points will be
linearly spaced over the selected Start/Stop frequency range, or
(correction points will be linearly spaced over the full frequency
range of the instrument). Use the @) (JJ keys to make your selection,
then press (Hz/s/ENTER).