Calibration Routine Errors - HP 83751A User Manual

Synthesized sweepers
Hide thumbs Also See for 83751A:
Table of Contents

Advertisement

606
"Warning! Learn String FW Revision Not Matched(606)"
The learn string that was received does not match the current
firmware revision. It was rejected and not used.
607
-200, "Execution error;Execution Not Allowed. Currently In
Restricted Mode(607)"
The instrument is in a restricted mode due to either the operation
of self test, or current operation of a calibration. Running of most
commands is not allowed. For best results, a device clear followed
by a *rst command should be sent.
608
-200, "Execution error;Execution Not Allowed. Currently In
Network Analyzer mode(608)"
Frequency zero can only be executed when the instrument is
in stand alone mode. When the instrument is connected to an
language, frequency zero cannot be implemented. If the user
attempts to implement frequency zero in these modes, an error
message will be generated.
-300, "Device specific error;Peaking Failed(701)"
701
For unspecified reasons, the CW peaking algorithms failed.
702
-300, "Device specific error;Peaking Never Leveled(702)"
ALC could not achieved level power under the current conditions.
703
-300, "Device specific error;Instrument Not in CW Mode(703)"
It is required that the instrument first be in CW mode before a YTF
peak is executed.
Error Messages
16-33