If The Maximum Leveled Power Check Fails - HP 83751A User Manual

Synthesized sweepers
Hide thumbs Also See for 83751A:
Table of Contents

Advertisement

Make sure that the RF OUTPUT connector was terminated before the peak
q
power-tracking sequence was run, and before the maximum leveled power
check.
Make sure the power you entered in steps 2 and 5 of the maximum leveled
q
power check procedure are correct for your particular instrument.
q
until the value measured equals the specified power. Note the sweeper's
front panel reading and use this value to repeat the maximum leveled
power check.
If the check still fails, contact the nearest Hewlett-Packard office for
q
assistance.
N O T E
If you need to ship your sweeper, remove the front handles (if so equipped) and use the original
packaging (or comparablel.
Performing the Operator's Check