-214
"Trigger deadlock"
Indicates that the trigger source for the initiation of a measurement
is set to GET and subsequent measurement query is received. The
measurement cannot be started until a GET is received, but the
GET would cause an INTERRUPTED error.
-215
"Arm deadlock"
Indicates that the arm source for the initiation of a measurement is
set to GET and subsequent measurement query is received. The
measurement cannot be started until a GET is received, but the
GET would cause an INTERRUPTED error.
-220
"Parameter error"
Indicates that a program data element related error occurred. This
error message 0 should be used when the device cannot detect the
more specific errors described for errors -221 through
-229.
-221
"Settings contlict"
Indicates that a legal program data element was parsed but could
not be executed due to the current device state. (See IEEE 488.2,
6.4.5.3 and 11.5.1.1.5).
-222
"Data out of range"
Indicates that a legal program data element was parsed but could
not be executed because the interpreted value was outside the
legal range as deEned by the device. (See IEEE 488.2, 11.5.1.1.5).
-223
"Too much data"
Indicates that a legal program data element of block, expression,
or string type was received that contained more data than the
device could handle due to memory or related device-speciEc
requirements.
Illegal parameter value
-224
Used where exact value, from a list of possibilities, was expected.
-225
Out of memory.
The device has insufficient memory to perform the requested
operation.
Error Messages