21- Full Selftst - HP 83751A User Manual

Synthesized sweepers
Hide thumbs Also See for 83751A:
Table of Contents

Advertisement

Special Functions
This function displays the results of the last time the self-test was run, or
displays
the test has not been run since the last time power was
turned on.
Pressing the
/
/
HZ
S
ENTER
must be disconnected prior to running the self-test. While the test is being
performed a **+ wait. +++ message will appear in the right-hand display,
and when the test is done, the results will be displayed. If the self-test
passed, the results will read, Full Test. PHS::;ED. If there is a failure, the
test that failed will appear in the MARKER/SWEEP/STATUS display, preceded
by -->. If more than one test failed, the most significant failure will be
displayed.
Before sweeper operation can be continued, a preset must be performed.
If the self-test fails (does not say
"Troubleshooting" chapter of the Service
*TST
DIAGnostic:TEST:FULLtest?
DIAGnostic:TEST:FULLtest:REPort?
15-28
key will activate the self-test. All external cables
Guide.
Associated SCPI Commands
Performs a fu/ self-rest
Executes the full self-test.
the most probable fat/ure test in the

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

83752b83752

Table of Contents