How Do I Use The Buffer - Keithley SourceMeter 2611 User Manual

Hide thumbs Also See for SourceMeter 2611:
Table of Contents

Advertisement

Series 2600 System SourceMeters User's Manual

How do I use the buffer?

Reference
The SourceMeter has two buffers per channel that can store from 1 to more than
100,000 readings. Each buffer reading is numbered and can also include the
source value and a timestamp.
The following example shows how to store 100 readings, source values, and
timestamps in Channel A, Buffer 1 and recall them from the front panel.
Step 1: Connect the DUT
Connect a 10kΩ resistor to the Channel A HI and LO terminals (Refer to
Figure
Step 2: Set up source and measure functions
Using the procedure described in
section, set-up source and measure functions:
Step 3: Configure the buffer
1.
2.
3.
4.
5.
6.
7.
8.
Refer to Section 7 of the Series 2600 Reference Manual for more
detailed information on using the buffer.
1-3).
Source function: volts
Source range: 20V (2611/2612) or 40V (2601/2602)
Source value: 10V
Measure function: current
Measure range: auto
Press the CONFIG key followed by the STORE key.
Choose COUNT, then set the number of readings to store to 100 using the
Rotary Knob, and press ENTER or the Rotary Knob.
Select CHANA_BUFF, then press ENTER or the Rotary Knob to select
Channel A.
Choose DEST, then CHANA_BUFFER1, and press ENTER or the Rotary
Knob.
Choose BUFFER1, then press ENTER or the Rotary Knob to set up Buffer 1.
Select CLEAR, then YES to clear the buffer.
Choose ELEMENTS, then enable (ON) both SRC-VAL (source value) and
TSTAMP (timestamp) storage.
Press EXIT several times to return to normal display.
Return to
"How do I make
measurements?" earlier in this
In this section:
2600S-900-01 Rev. A / May 2006
Front Panel Operation 1-15

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sourcemeter 2602Sourcemeter 2612Sourcemeter 2601