How do I use the buffer?

Reference
The instrument has two buffers per channel that can store from 1 to more than 100,000 readings.
Each buffer reading is numbered and can also include the source value and a timestamp.
The following example shows how to store 100 readings, source values, and timestamps in
Channel A, Buffer 1 and recall them from the front panel.
Step 1: Connect the DUT
Connect a 10kΩ resistor to the Channel A HI and LO terminals (see
Step 2: Set up source and measure functions
Using the procedure described in
and measure functions:
Source function: volts
Source range: 20V (2611/2612/2635/2636) or 40V (2601/2602)
Source value: 10V
Measure function: current
Measure range: auto
Step 3: Configure the buffer
1.
From the front panel, press the STORE key.
The Count screen displays.
2.
Use the navigation wheel to select the number of readings and then press ENTER.
NOTE You can store up to 100.
3.
Choose CONFIG > STORE > CHANA_BUFF, then the press the navigation wheel to select
Channel A.
4.
Select DEST > CHANA_BUFFER1, and then press ENTER.
NOTE The message "Cant' change dest buffers" while storing displays if the
5.
Choose CHANA-BUFF1, then press the navigation wheel to set up Buffer 1.
Complete the following steps to clear the Buffer 1.
1.
Select BUFFER1 > CLEAR.
2.
Select YES.
3.
Choose ELEMENTS.
4.
Select SRC-VAL and then choose ON to enable the source value.
5.
Select TSTAMP and then choose ON to enable timestamps.
6.
Press EXIT until the normal display.
1-16
See Section 7 of the Series 2600 Reference Manual for more detailed information on
using the buffer.
"How do I make measurements?" on page
instrument is in store mode.
Return to
Section Topics
Series 2600 System SourceMeter
Figure 1-4 on page
2600S-900-01 Rev. C / January 2008
®
Instruments User's Manual
1-14).
1-12, set up source

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

26022612261126352636

Table of Contents