How Do I Store Measurements In Non-Volatile Memory; Front Panel Operation; Remote Programming - Keithley SourceMeter 2611 User Manual

Hide thumbs Also See for SourceMeter 2611:
Table of Contents

Advertisement

Series 2600 System SourceMeters User's Manual
How do I store measurements in non-volatile
memory?
A Series 2600 SourceMeter has two non-volatile buffers for measured readings,
source values, and timestamps: NV Buffer 1 ( nvbuffer1 ) and NV Buffer 2
( nvbuffer2 ). Data stored in these buffers will not be lost when the Series 2600 is
turned off.
Reference

Front panel operation

Refer to
ing data from the front panel of the Series 2600.

Remote programming

The following example demonstrates a typical way to use a reading buffer. The
commands in the following steps will perform three SMU A voltage
measurements, store them in nvbuffer1 , and then read the buffer:
Step 1: Clear and configure the buffer
The following command sequence will clear the buffer of all readings, and
configure the buffer to store timestamps and source values:
smua.nvbuffer1.clear()
smua.nvbuffer1.collecttimestamps = 1
smua.nvbuffer1.collectsourcevalues = 1
NOTE
Step 2: Perform measurements
The following command sequence sets the measure count to three, turns on the
output, performs the three measurements, and then turns off the output:
smua.measure.count = 3
smua.source.output = smua.OUTPUT_ON
smua.measure.v(smua.nvbuffer1)
smua.source.output = smua.OUTPUT_OFF
Refer to Section 7 of the Series 2600 Reference Manual for details
on the buffer (data store).
"How do I use the
buffer?" in
Setting collecttimestamps to 0 and collectsourcevalues
to 0 will disable the storage of timestamps and source
values.
Return to
Frequently Asked Questions
Section 1
of this manual for details on stor-
In this Appendix:
2600S-900-01 Rev. A / May 2006