1-16
Front Panel Operation
Step 4: Turn on the output
Press the CHAN A (Models 2602/2612) OUTPUT ON/OFF key to turn on the
source output.
Step 5: Store readings
1.
2.
Step 6: Turn off the output
Press the OUTPUT ON/OFF key to turn off the output after storage has
completed.
Step 7: Recall readings
1.
2.
3.
4.
2600S-900-01 Rev. A / May 2006
Press the STORE key to store readings. The asterisk (*) annunciator turns
on to indicate data storage operation is enabled.
Press EXIT to stop data storage before it finishes.
Press the RECALL key to access buffer readings (repeatedly pressing
RECALL will cycle through Buffer 1 then Buffer 2 for Channel A and then
Channel B). A message will be displayed if a buffer is empty.
Note the buffer display data (Refer to
The reading is on the top display at the left.
The buffer location number is on the right. For example, location
#000001 indicates that the displayed reading is stored at the first
memory location.
The source value is positioned at the lower left side of the display.
The timestamp is positioned at the lower right side. The first source-
measure reading stored in the buffer (#000001) is timestamped at
00000000.001 seconds. Subsequent readings are timestamped
relative to when the time storage was started. The interval between
readings will depend on the reading rate.
To display the other readings stored in the buffer, choose the desired
memory location number:
Use the Rotary Knob to increment and decrement the selected digit of
the location number.
Set the cursor position with the Rotary Knob or CURSOR keys.
To exit from the data store recall mode, press EXIT.