Series 2600 System SourceMeters User's Manual
Table 4-1
Assigning a node number to an instrument from the front panel
1) Press the MENU key to access MAIN MENU.
2) Select COMMUNICATION menu.
3) Select TSPLINK_CFG menu.
4) Select NODE menu.
5) Set the NODE number (01 to 64) for the instrument.
For remote programming, the tsplink.node attribute is used to set the node
number for an instrument:
tsplink.node = N
where : N = 1 to 64
The node number of an instrument can be determined by reading the
tsplink.node
nodenumber = tsplink.node
print(nodenumber)
The above
number is 1, the value
Step 3: Reset the TSP-Link
Reference
After all the node numbers are set, you must reset (initialize) the TSP-Link
interface system. For initialization to succeed, all units must be powered on when
the TSP-Link reset is performed.
The TSP-Link reset can be performed from the MAIN MENU of any instrument in
the system; this process is summarized in
Table 4-2
Resetting the TSP-Link from the front panel
1) Press the MENU key to access MAIN MENU.
2) Select COMMUNICATION menu.
3) Select TSPLINK-CFG menu.
4) Select RESET to initialize the TSP-Link.
For remote programming, the following command initializes the TSP-Link:
tsplink.reset()
Series 2600 MAIN MENU
attribute as follows:
command will return the node number. For example, if the node
print
1.000000e00
Refer to "Initialization" in Section 9 of the Series 2600 Reference
Manual for more information on resetting the TSP-Link.
Series 2600 MAIN MENU
Return to
In this section:
Controlling Multiple SourceMeters 4-3
will be returned.
Table
4-2.
2600S-900-01 Rev. A / May 2006