SENSOR LIMITATIONS
FCC COMPLIANCE
This device has been tested and found to comply with the limits for a Class A digital device, pursuant to
part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment. This equipment generates,
uses, and can radiate radio frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the user will be
required to correct the interference at user's own expense.
Operation is subject to the following two conditions: (1) This device may not cause harmful interference,
and (2) this device must accept any interference received, including interference that may cause
undesired operation.
Shop for Dent products online at:
USB port
USE ONLY SHUNTED CURRENT TRANSFORMERS (CTs).
Do not use other CTs. Only use shunted CTs with a 333mV maximum output only.
Serious shock hazard and logger damage can occur if unshunted CTs are used. The
UL listing covers the use of the following DENT Instruments CTs that are UL
Recognized and have been evaluated to IEC 61010-1:
CT-HSC-020-X (20A Mini), CT-HSC-050-X (50A Mini), CT-HMC-0100-X (100A Midi),
and the CT-HMC-0200-X (200A Midi). The use of any other CT will invalidate the UL
Listing of the ELITEpro XC.
Pulse: Use only "dry contact" non-energized pulse inputs (ELITEpro SP only). Use
of energized pulse initiators can cause damage to the logger and a potential shock
hazard to the user. Wiring must meet 600V AC CAT III rating.
Equipment protected throughout by double insulation (IEC 536 Class II). CAT III 80-
600 VAC 125mA 50/60 Hz
PowerMeterStore
www.
9
1.877.766.5412
.com