NOTE
Emerson Process Management recommends 0.4% O
calibration gases.
O2 Reset Values - Resets factory default values.
O2 Cal Interval - If automatic calibration is selected, this selects the interval
between calibrations.
O2 Next Cal - If automatic calibration is selected, this selects the time until
the first initial calibration takes place.
Gas Time - How long should each cal gas flow. Factory default is 300
seconds, but the user may want to vary this depending on the length of
calibration gas tubing runs.
Purge Time - Used if the O
calibration. After the second cal gas is removed, how long until the sensor
comes back to the normal process reading, and the fieldbus digital signal can
be released.
Auto Calib? - Select "Yes" if an SPS or IMPS autocalibration system is part of
the system.
SYSTEM/Input/Output
Analog
Pertaining to the analog fieldbus digital signal representing O
O2 Type - Fieldbus digital signal may be configured to increase with
increasing O
or the reverse.
2
O2 Range - Upper O
O2 Alarm Level - User can configure the digital output to alarm at a given
O
level.
2
Digital
A bi-directional logic signal may be configured as an alarm, or as a calibration
handshake signal.
Logic I/O Mode - One of 9 different sets of conditions can be set for the
digital signal. See Table 8-2.
Low O2 Alarm - If any of the conditions noted above include a low O
process alarm, set the value here.
Input State - Notes the current condition of the bi-directional digital signal.
Force Output - Forces the output state of the signal to either open or
closed. This is used primarily when diagnosing potential problems with this
signal.
SYSTEM/Parameters
O2 Slope - O
slope is data regarding the strength of the sensing cell output.
2
This information is automatically calculated after a calibration, and the user
does not normally input this data.
output is selected to hold the last value during
2
range is user selectable.
2
Oxymitter 5000
and 8% O
for
2
2
.
2
2
6-7