Oxymitter 5000 Configuration - Emerson Oxymitter 5000 Instruction Manual

Oxygen transmitter
Hide thumbs Also See for Oxymitter 5000:
Table of Contents

Advertisement

Oxymitter 5000
Figure 4-1. Electronics Housing
Terminals with LOI
Fieldbus Digital
Signal
Logic I/O
Oxymitter 5000
Configuration
4-2
Terminal
Block
Ground Lugs
Remove power before changing defaults. If defaults are changed under power, damage to
the electronics package may occur.
Located on the microprocessor board, the top board, is a switch that controls
the simulate enable status of the Oxymitter 5000 (Figure 4-2). To access this
switch, the LOI module must be removed. To allow the Oxymitter to be placed
in simulation mode, place position two of SW2 in the ON position. Once the
Oxymitter has been set to simulate mode, switch position two of SW2 to the
OFF position to remove the Oxymitter from simulate mode. Note that SW2
does not actually place the Oxymitter in simulate mode, it only allows the
Oxymitter to be placed into simulate mode through the fieldbus interface.
Positions 1, 3 and 4 of SW2 are not used, and should remain in the OFF
position.
Typically, the probe's sensing cell, in direct contact with the process gases, is heated to
approximately 1357° F (736° C). The external temperat ure of the probe body may exceed
842°F (450° C). If operating conditions also contain high oxygen levels and combustible
gases, the Oxymitter 5000 may self-ignite.
Rosemount Analytical Inc.
SMART FAMILY
R
Solon, OH 44139
TM
HART
800-433-6076
OXYMITTER 5000
TM
SERIAL NO.
TAG NO.
VOLTS:
WATTS:
500 VA
85-264 VAC 48-62 Hz
OUTPUT:
4-20 mA
LINE
FUSE:
5 Amps
Instruction Manual
IM-106-350, Rev 2.2
Oxymitter 5000
Electronics
Housing
LOI
July 2008