Emerson Oxymitter 5000 Instruction Manual page 188

Oxygen transmitter
Hide thumbs Also See for Oxymitter 5000:
Table of Contents

Advertisement

Oxymitter 5000
Parameter Mnemonic
SENSOR_TYPE
ST_REV
STATS_ATTEMPTS
STATS_FAILURES
STATS_TIMEOUTS
STRATEGY
TAG_DESC
TIME_TO_NEXT_CAL
TRANSDUCER_
DIRECTORY
TRANSDUCER_TYPE
UPDATE_EVT
VERSION
XD_ERROR
Table C-1. Calibration State
Values
Code Value
Description
0
Normal System Operation
1
Calibration Required
2
Apply Test Gas 1
3
Test Gas 1 Flow
4
Test Gas 1 Read
5
Test Gas 1 Done
6
Apply Test Gas 2
7
Test Gas 2 Flow
8
Test Gas 2 Read
9
Test Gas 2 Done
10
Abort/Fail
11
Stop Gas
12
Purge
C-4
Initial
Valid Range
Value
0.0-9999.0
External Event Required to
Units
Selected from list in Transducer Block
Specification, part 2. FF-903, page 40,
section 4.3.
See Function Block Specification, part 1.
FF-890, page 49.
This shows the number of communication
attempts between the Oxymitter and the
internal fieldbus interface card.
This shows the number of communication
failures between the Oxymitter and the
internal fieldbus interface card.
This shows the number of communication
failures due to reply time-out between the
Oxymitter and the internal fieldbus interface
card.
See Function Block Specification, part 1.
FF-890, page 49.
See Function Block Specification, part 1.
FF-890, page 49.
Hours
This is the time remaining until the next
automatic calibration.
See Transducer Block Specification, part 1.
FF-902, page 11.
Selected from list in Transducer Block
Specification, part 2. FF-903, page 39,
section 4.2.
See Function Block Specification, part 2.
FF-891, page 45.
N/A
This is the version of the Oxymitter software.
See Transducer Block Specification, part 2.
FF-903, page 38.
Go to Next Step?
Yes
Yes
Yes
No
No
No
Yes
No
No
No
Yes
Yes
No
Instruction Manual
IM-106-350, Rev 2.2
Description
July 2008
Parameter
Number
31
1
56
57
58
3
2
48
9
10
7
59
11