Oxymitter 5000
Figure 1-5. Oxymitter 5000
FOUNDATION Fieldbus
Connections
with Integral Electronics
2 Calibration Gas Lines
by Customer
[
300 ft 90 m
(
1-8
A source of instrument air is optional at the Oxymitter 5000 for reference
air use. Since the unit is equipped with an in place calibration feature,
provisions can be made to permanently connect calibration gas bottles to the
Oxymitter 5000.
If the calibration gas bottles will be permanently connected, a check valve is
required next to the calibration fittings on the integral electronics.
This check valve is to prevent breathing of the calibration gas line and
subsequent flue gas condensation and corrosion. The check valve is in
addition to the stop valve in the calibration gas kit or the solenoid valves in the
IMPS 4000 or SPS 4001B.
NOTE:
The electronics is rated NEMA 4X (IP66) and is capable of operation at
temperatures up to 185° F (85° C).
The optional LOI is also rated for operation at temperatures up to 185°F
(85° C). The infrared keypad functionality will degr ade at temperatures above
158° F (70°C).
Retain the packaging in which the Oxymitter 5000 arrived from the factory in
case any components are to be shipped to another site. This packaging has
been designed to protect the product.
Oxymitter 5000
) max]
Line Voltage
Fieldbus Digital
Signal
Instruction Manual
IM-106-350, Rev 2.2
Fieldbus Computer
Terminal
July 2008