Figure 9-8. Heater Strut
Assembly
V-Deflector
Ceramic
Diffuser
Cell Flange
Assembly
9-16
Ceramic Support Rod
Heater
13. For units with remote electronics, install the probe head as follows:
a. See Figure 9-9. Make sure that the O-ring is in good condition. Seat
the O-ring in the mating groove of the probe.
b. Insert the probe signal cables into the housing.
c. Turn the conduit ports of the housing to the CAL and REF gas ports
side of the probe and position the housing on the probe.
d. Install and tighten four screws (2, Figure 9-4).
e. Reconnect the probe signal cables to the probe signal and heater
wire connectors, Figure 9-9. Make sure the connectors are secure.
f. Install and tighten cover.
14. Follow the instructions in "Removal and Replacement of Probe" to
install the Oxymitter 5000 into the stack or duct.
Instruction Manual
IM-106-350, Rev 2.2
July 2008