Laying of cables
Cable for use in Zone 1 and 2 or 21 and 22 must satisfy the requirements for having a proof
voltage > 500 V AC applied between the conductor/ground, conductor/shield and
shield/ground.
Connect the devices that are operated in hazardous areas as per the stipulations applicable
in the country of operation, e.g. for Ex "d" and "nA", permanent cables must be laid
With intrinsically safe circuits, use only certified meters appropriate for the transmitter.
If a non-conforming supply unit is used, the "fail-safe" type of protection will no longer be
effective and the approval certification will be invalid.
2.3.1

ATEX 2 GD approval

The device is approved for use in hazardous area and has certificate Sira 11ATEX2124X.
Remote transmitter version:
II 2(1) GD
EX d e [ia] ia IIC T6 Gb
Ex tD A21 IP67 T85 °C
Ta -25 °C to 60 °C
Note
Markings for remote version only
These markings are for the Remote version only; the marking applicable to the Compact
versions depend upon their construction. See the latest issues of certificates Sira
07ATEX3181X, Sira 07ATEX1182X and Sira 03ATEX3339X.
MAG 6000 I / MAG 6000 I Ex de transmitter
Operating Instructions, 04/2016, A5E02083319-AG
WARNING
WARNING
Safety notes
2.3 Installation in hazardous area
13