Alarm, error, and system messages
8.2 List of error numbers
Error text
Error
No.
Remedy text
42
P42 - Current output
Check cables
P43 - Internal error
43
Switch off and on
P44 - CT SENSORPROM®
44
F60 - CAN comm. error
60
Transmitter/AOM
F61 - SENSORPROM® error
61
Replace
F62 - SENSORPROM® ID
62
Replace
F63 - SENSORPROM®
63
Replace
F70 - Coil current
70
Check cables
F71 - Internal error
71
Replace transmitter
54
Comment
Current loop is disconnected or the loop resistance
is too big
Too many errors occurred at the same time.
Some errors are not detected correctly
SENSORPROM
unit has been used as CT version
®
CAN bus communication error. An add-on module,
the display module or the transmitter is defective
It is not possible to rely on the data in
SENSORPROM
unit anymore
®
The SENSORPROM
unit ID does not comply with
®
the product ID.
The SENSORPROM
unit is from another type of
®
product SITRANS F C, SITRANS F US etc.
It is not possible to read from the SENSORPROM
unit anymore
Coil excitation has failed
Internal conversion error in ASIC
Output
status
Active
Active
Active
Zero
Active
Zero
Active
®
Active
Active
MAG 6000 I / MAG 6000 I Ex de transmitter
Operating Instructions, 04/2016, A5E02083319-AG
Input
status
Active
Active
Active
Inactive
Active
Inactive
Active
Active
Active