9.2 Calibration of HYPROP BALANCE Series

Applicable models: HYPROP BALANCE
※ Span adjustment should be performed with the balance installed perfectly level and without any
load on the measurement pan.
1
Invoke the span-adjustment function.
Hold down.
2
Start span adjustment.
3
Zero-point calibration
4
Calibration of weighing capacity point
Calibration
Hold down the
release your finger when the display
switches from [Func] to [S.A. CAL].
Release your
finger.
Press the
down the
release both fingers simultaneously.
Display will change to a flashing [CAL.
0] and the zero-point will be calibrated
automatically.
Release both
fingers at the
same time.
Once zero-point calibration is
completed, display switches to [CAL.
on]. Calibration now proceeds to that
of the weighing capacity point.
Slowly and fully turn the calibration
knob in the direction of [Calibration].
The balance assumes the state where
the calibration weight is loaded on the
measurement pan.
The display starts flashing and the
balance automatically adjusts the
weighing capacity point.
Function
key, then
Function
key while holding
Zero/Tare
key, then
29