The percentage measurement function operates by weighing an actual sample, selected as the
reference, and saving its weight as the reference value and indexing it as 100%. When a
measurement sample loaded on the balance is lighter or heavier than the reference, its weight is
indicated as a percentage (%) value relative to the reference weight.
1
Activating the percentage mode
2
Setting the reference value
3
Loading the sample
4
Saving the reference value
5
Loading a sample to be measured
20

7. Measuring Percentage

Press the
Function
[%].
Press the
Set
key.
The display begins flashing
[P.
SEt ]. The balance is now ready
for reference value setup.
Load the reference sample.
Press the
Memory
The balance indexes the weight value
of the reference sample as 100% and
returns you to measurement mode.
The balance now indicates the weight of
the loaded sample as a percentage (%)
value relative to the reference value.