Limit Function; Limit Function Setup - UMS HYPROP Operation Manual

High-precision tuning fork electronic balance
Hide thumbs Also See for HYPROP:
Table of Contents

Advertisement

The limit function judges measurements according to a limit value saved in the balance.
The function shows the judgement result by displaying the [
(appropriate), or LO (insufficient). This function is very useful when discriminating between
conforming and nonconforming articles. It is also useful when measuring a given constant quantity
consecutively, in conjunction with a range of reference weights defined by upper- and lower-limit
values.
This function can be used in weight mode, counting mode, or percentage mode.
Limit value input methods
Either of the following two methods can be used in the different modes:
(1)
Actual quantity setup method ........ An actual sample is loaded on the balance and its weight
(2)
Numeric value setup method ......... The limit value is entered with a key stroke.
※The limit values entered are held in memory, even when the balance is powered down.
※The respective limit values for weight mode, counting mode, and percentage mode are set up
independently.
Indication of judgement result
The [
] mark lights up as either HI, OK, or LO on the left side of the display, indicating the result of
judgement.
Judgement Results
HI (excessive)
OK (appropriate)
LO (insufficient)

8.1 Limit Function Setup

1
Invoking the function
Continuous
pressing
22

8. Limit Function

saved as the limit value.
Upper/lower-limit setting
Upper-limit value < measurement
value
Upper-limit value  measurement
value  lower-limit value
Lower-limit value > Measurement
value
] mark as either HI (excessive), OK
No indication
Limit value  measurement value
Limit value > Measurement value
Press and hold down the
key.
Release the key when [Func] is
displayed.
The display changes to the function
Key
setup and the first item is displayed.
released
One-point setting
Function