UMS HYPROP Operation Manual page 24

High-precision tuning fork electronic balance
Hide thumbs Also See for HYPROP:
Table of Contents

Advertisement

☆ Key Points of the Procedure ☆
1.
While samples are being saved, the value indication disappears temporarily, and only the [M]
mark flashes. If the balance is affected by wind or other vibrations during this process, the saving
time may be prolonged.
2.
If [L-Err] appears briefly, it indicates one of the following states:
(1)
The weight of the reference sample is insufficient.
For the limit weight that can be saved (% limit weight), please refer to "13. Specifications,"
on page 41.
(2)
While setting up the reference value in Step 2, the
samples were loaded on the balance.
※If [L-Err] appears, sampling has been interrupted and the sample value being processed will
not be saved.
3.
The minimum intervals between percentages in the unit switch from 1%, to 0.1%, to 0.01%,
depending on the reference weight from the sampling.
Measurable unit weight in counting and percentage mode
Model
Measurable unit weight in
counting mode [g]
Minimum weight in
percentage mode [g]
Set key has been pressed while the
HYPROP BALANCE-2200CE
0.01
1
21