Since electronic balances are affected by gravity gravitational acceleration, they produce different
values in various locations. Therefore, before use, balances must be calibrated at the location where
they are installed. Calibration is also required after long periods without use, or if a balance begins to
produce inaccurate values.
Calibration of a balance, or "span adjustment," is required to produce accurate measurements.

9.1 Calibration of HYPROP BALANCE Series

Applicable models: HYPROP BALANCE
※ Span adjustment should be performed with the balance installed perfectly level and without any
load on the measurement pan.
1
Activating the span adjustment function
Pressed
continuously
2
Starting the span adjustment
3
Zero calibration

9. Calibrating the Balance

Key released
Release
simultaneously.
Press and hold down the
key.
Release the key when the display
changes from [Func] to [CAL].
Press the
Function
down the
Zero/Tare
release both keys simultaneously. The
0 ], indicating that
display flashes [on
zero-point calibration is underway.
Zero-point calibration is finished when
the display changes to [on
Proceed to the calibration of the
capacity point.
Function
key while holding
key, then
F.S ].
27