Figure 32
The Average acquisition mode decreases the screen refresh rate.
To select the Average acquisition mode:
1 Press [Acquire].
2 In the Acquire menu, press Acquisition.
3 Continue pressing the Acquisition softkey or turn the
4 Press Averages and turn the

To select the Peak Detect acquisition mode

In Normal or Average acquisition modes, at longer horizontal time/div
settings, the oscilloscope's analog- to- digital converter samples at a rate
that yields more samples than can be stored in a limited amount of
oscilloscope memory. Consequently, samples are thrown away (decimated),
and you can miss narrow excursions on a signal.
Agilent 1000B Series Oscilloscopes User's Guide
Noisy Waveform With Averaging
select "Average".
(2, 4, 8, 16, 32, 64, 128, or 256).
entry knob to select the desired number
Capturing Data
3
entry knob to
77