Download Print this page
Agilent Technologies 16048D Operation And Service Manual
Agilent Technologies 16048D Operation And Service Manual

Agilent Technologies 16048D Operation And Service Manual

Test leads

Advertisement

Quick Links

sales@artisantg.com
artisantg.com
(217) 352-9330 |
|
Click HERE
Find the Keysight / Agilent 16048A at our website:

Advertisement

loading

Summary of Contents for Agilent Technologies 16048D

  • Page 1 sales@artisantg.com artisantg.com (217) 352-9330 | Click HERE Find the Keysight / Agilent 16048A at our website:...
  • Page 2 Agilent 16048D Test Leads Operation and Service Manual Fourth Edition Agilent Technologies Japan, Ltd. Agilent PN 16048-90031 May 2000 Printed in Japan Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 3 In addition it violates safety standards of design, manufacture, and intended use of the instrument. The Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements. 16048D comply with INSTALLATION CATEGORY I and POLLUTION NOTE DEGREE 2 in IEC61010-1.
  • Page 4 Because of the danger of introducing additional hazards, do not install substitute parts or perform unauthorized modifications to the instrument. Return the instrument to a Agilent Technologies Sales and Service Office for service and repair to ensure that safety features are maintained.
  • Page 5 Instruments are warranted for a period of one year. Fixtures and adapters are warranted for a period of 90 days. During the warranty period, Agilent Technologies will, at its option, either repair or replace products that prove to be defective.
  • Page 6 Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For any assistance, contact your nearest Agilent Technologies Sales and Service Office. Addresses are provided at the back of this manual. Safety Symbol General definitions of safety symbols used on the instrument or in manuals are listed below.
  • Page 7 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 8: Table Of Contents

    Contents 1. Operation Product Description ..........10 Specifications .
  • Page 9 Contents Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 10: Operation

    Operation This operating note provides complete information on the 16048D Test Leads. The 16048D is shown pictorially in Figure 1-1, its physical dimensions are given in Table 1-1. To order additional copies of this operating note, use the part number listed on the rear cover.
  • Page 11: Product Description

    BNC (m) terminated-coaxial test leads. These test leads are used to attach user-fabricated test fixtures. DC bias levels of up to ±40V can be applied to the 16048D. Cable length is 2 meter. The 16048D is shown in Figure 1-1.
  • Page 12: Specifications

    Operation Specifications Specifications Table 1-1 Specifications of the 16048D Function: 4-terminal pair type fixture which is equipped with four BNC (m) terminated-coaxial test leads. Connector Type: BNC male Maximum Voltage: ± 40 V peak max. (AC+DC) Cable Length: Weight: 460 g...
  • Page 13: Compensation For Fixture Residual Impedance Error

    Compensation for Fixture Residual Impedance Error Compensation for Fixture Residual Impedance Error The 16048D has inherent stray capacitance, residual inductance, and residual resistance that affect the accuracy of measured values. To compensate for, or negate, these residuals to minimize measurement error, the instrument’s Open/Short compensation procedure should be performed.
  • Page 14: Operation

    Operation Setup and measurement procedure is as follows: 1. Connect the 16048D directly to the UNKNOWN terminals of the instrument. 2. Connect the user-fabricated test fixture to the test leads. 3. Perform Open/Short compensation as described in the instrument’s operation manual.
  • Page 15 Operation Operation Chapter 1 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 16: Service

    Service Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 17: Maintenance

    Service Maintenance Maintenance An exploded view of the 16048D (for parts identification) is shown in Figure 2-1. Do not disassemble any further than shown. Maintenance consists principally of cleaning contacts and replacing worn or damaged parts. Take special care when cleaning contacts.
  • Page 18 Service Maintenance Figure 2-1 Parts Identification Chapter 2 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 19 Service Maintenance Table 2-1 Parts Identification Agilent Part Reference Qty. Description SLEEVE-METAL NUT-HEX WASHER-FL MTLC WASHER-FL NM 0400-0203 GROMMET-RND 16048-61612 WIRE ASSEMBLY BOTTOM COVER INSULATOR CONNECTOR-BNC BNC-ASSEMBLY 2200-0103 SCREW-MACH 4-40 2190-0206 WASHER-FL MTLC 16047-40000 STOPPER 16048-04011 TOP COVER 2360-0192 SCREW-MACH 6-32 1400-0719 CABLE TIE GROMMET-RND...
  • Page 20 Service Maintenance Agilent Part Reference Qty. Description TEST LEAD (1 thru 29) 16048-60020 1250-0118 CONNECTOR-RF BNC 2360-0115 SCREW-MACH 6-32 16032-10021 PLATE 2190-0016 WASHER-LK INTL T 2950-0001 NUT-HEX 16032-10022 PLATE SHIELD 16032-60001 BNC BRACKET (30 thru 35) a. Agilent internal-only part number. *: Not separately replaceable.
  • Page 21 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 22 REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative.

This manual is also suitable for:

16048a