Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...
Page 2
Agilent 16034E Test Fixture Operation and Service Manual Third Edition Agilent PN 16034-90041 December 1999 Printed in: Japan Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 3
WARNINGS elsewhere in this manual may impair the protection provided by the equipment. In addition it violates safety standards of design, manufacture, and intended use of the instrument. The Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements. • Ground The Instrument...
Page 4
Because of the danger of introducing additional hazards, do not install substitute parts or perform unauthorized modifications to the instrument. Return the instrument to a Agilent Technologies Sales and Service Office for service and repair to ensure that safety features are maintained.
Page 5
Instruments are warranted for a period of one year. During the warranty period, Agilent Technologies will, at its option, either repair or replace products that prove to be defective.
Page 6
Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For any assistance, contact your nearest Agilent Technologies Sales and Service Office. Addresses are provided at the back of this manual. Safety Symbol General definitions of safety symbols used on the instrument or in manuals are listed below.
Operation This Operating note provides Operating and service information for the Agilent 16034E Test Fixture. Specifications are listed in Table 1-1. Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
This test fixture is designed to facilitate easy and accurate measurements of chip type components and similar types not furnished with lead wires. Connection of sample components to the 16034E is done by simply pressing the loading lever to left, inserting sample components in the V-mount center section (between fixture jaws) and releasing lever.
Operation The set up and measurement procedure is as follows: 1. Set the cable length to 0m. 2. Connect the 16034E directly to the UNKNOWN terminals. 3. Loosen the knob A. Figure 1-2 4. Adjust the contact pin position on the V-mount center section with the knob B to adapt your DUT size.
Page 13
Operation Operation 10. Perform the open compensation described in the instrument’s operation manual. 11. Inserting a sample components in the V-mount center section (between fixture jaws). 12. Releasing the lever C to fix the DUT. Figure 1-4 WARNING DO NOT APPLY MORE THAN ± 40V OF DC BIAS VOLTAGE TO THE UNKNOWN TERMINALS.
Operation Specifications Specifications Table 1-1 Specification of the 16034E. Function: For use with 4 terminal-pair LCR Meters and Impedance Analyzers. Measurable components: Chip type components ( Not furnished with lead wires) Dimension of measurable components: ≤ 40 MHz Frequency: Maximum Voltage: ±...
Operation Specifications Typical Characteristics Table 1-2 Typical characteristics ≥ Incremental Error at f 1 MHz Parameter reading error (%) Offset Value for D Á ± × ----- - NOTE f is the measurement frequency in MHz. The incremental errors calculated from the equations given in this table are additive when measurements are made above 1MHz.
Maintenance Maintenance An exploded view of the 16034E for parts identification is shown in Figure 2-1 and Figure 2-2. Do not disassemble any further than shown. Maintenance consists principally of cleaning contacts and replacing worn or damaged parts. Take special care when cleaning contacts.
Page 18
Service Maintenance Figure 2-1 Parts Identification Chapter 2 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 19
Service Maintenance Figure 2-2 Parts Identification Table 2-1 Reference Agilent Part No. Qty. Description Designation 16034-00221 TOP COVER 2200-0103 SCREW 1460-0323 SPRING 16034-01202 BRACKET 16034-25004 KNOB 2360-0201 SCREW 3050-0228 WASHER 16034-05001 LEVER 16034-23001 SHAFT 16034-25002 COLLAR 0510-0083 ERING 3050-1138 WASHER 2420-0006 16034-20002 BLOCK...
Page 21
Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...