Phase Between Rising Edges

Measures the phase between two waveforms using the rising edges.
Phase is the calculated phase shift from source 1 to source 2, expressed in
degrees. Negative phase shift values indicate that the rising edge of source
1 occurred after the rising edge of source 2.
Time
origin
Figure 45

Phase Between Falling Edges

Measures the phase between two waveforms using the falling edges. See
Figure 45
Agilent 1000B Series Oscilloscopes User's Guide
Delay
------------------------------------- -
Phase
=
Source 1 Period
line
50%
Phase Measurements
on page 109.
×
360°
Period
Delay
50%
Phase from channel 1
to channel 2 rising edge
Making Measurements
Channel 1
Channel 2
4
109