Purge the pump and fill the line to be used for the test with water.
Zero the scale. Weigh the collection vessel on a calibrated scale or balance.
Take the collection vessel and go to the pump and set the pump flow for 0.9 mL/minute.
Start the pump. (Direct the initial flow of water off to waste until the pressure stabilizes.
When pressure stabilizes, direct the flow of water into the collection vessel and
simultaneously start a stopwatch or other timing device.
Let the pump run for 10minutes.
When 10minutes has elapsed, remove the tube from above the collection vessel and
insert it into the waste.
Weigh the collection vessel with the water. Subtract the weight of the collection vessel
The remaining weight of the water should be 9 grams ± 0 .09 g (8.91 to 9.09g)
0.9 mL/min * 10 min = 9 g (water weighs 1 gram per mL)
Divide the weight of the water by 10 to get the volume delivered in a 1 minute interval.
Record this number in the Test Result field of Flow Accuracy @ 0.9 mL/min portion of the
IQOQ Test Certificate. Enter N/A if testing Analytical Pump
2.4.5
Test Results
Parameter
FX-10 Pump (30 µ µ µ µ L
Piston) Accuracy: 0.9
mL/min
Comments:
PerkinElmer LAS Representative:
Reviewed By:
Flexar Analytical / FX-10 UHPLC Pump IQ/OQ Rev. D June 2009
Specification
± 1% of 0.9 mL/min
(0.891ml/min to
0.909ml/min)
Test Result
Pass/Fail
Date:
Date:
.
Comments