FL 6500/ 8500 W ater Jacketed 4-Position M ulti-Cell
Holder w / Stirrer I nstallation I nstructions
This instruction sheet describes the installation of this accessory which is used with the FL 6500/8500
Fluorescence Spectrometer.
Read these instructions before you install this accessory.
NOTE:
Contacting P erkinElm er
Supplies, replacement parts, and accessories can be ordered directly from PerkinElmer, using the part
numbers.
See our website:
http://perkinelmer.com
PerkinElmer's catalog service offers a full selection of high-quality supplies.
To place an order for supplies and many replacement parts, request a free catalog, or ask for information:
If you are located within the U.S., call toll free 1-800-762-4000, 8 a.m. to 8 p.m. EST. Your order will be
shipped promptly, usually within 24 hours.
If you are located outside of the U.S., call your local PerkinElmer sales or service office.
Features
• Controls the temperature by circulation of liquid
• 4-position cell holders
• Built-in stirring mechanism
• Control stirring speed using Spectrum FL software
Figure 1 Water Jacketed 4-Position Multi-Cell Holder w/Stirrer (P/N:N4201027)
PerkinElmer, 710 Bridgeport Avenue,
Shelton, CT 06484-4794, U.S.A
Produced in the USA.

Advertisement

loading

Summary of Contents for PerkinElmer FL 6500

  • Page 1 FL 6500/ 8500 W ater Jacketed 4-Position M ulti-Cell Holder w / Stirrer I nstallation I nstructions This instruction sheet describes the installation of this accessory which is used with the FL 6500/8500 Fluorescence Spectrometer. Read these instructions before you install this accessory.
  • Page 2: Dimensions And Specifications

    09931525A Dim ensions and Specifications Physical Characteristic Specifications Moving distance of one cell (mm) 12.8 Moving distance of four cells (mm) 38.4 Moving time of one cell (sec) 1.55 Moving time of four cells (sec) Dimensions (mm) 190 (W) X 267 (D) X 98 (H) Weight (Kg) 1.84 Environm ental Conditions for Operation...
  • Page 3: Installation

    Figure 3 Two Male connectors of quick coupling I nstallation Prepare the FL 6500/8500 Fluorescence Spectrometer to install this accessory. Connect the instrument power cord and the communication cable. Loosen the accessory fixing bolt and remove the current sample accessory.
  • Page 4 09931525A After checking the pogo pin position of the sample compartment, place the Water Jacketed 4- Position Multi-Cell Holder w/Stirrer into the pogo pin. Figure 7 Installing the accessory Fix the front cover plate of the Water Jacketed 4-Position Multi-Cell Holder w/Stirrer with the bolts.
  • Page 5 09931525A 10. Assemble the tube with the male connector of quick coupling. Figure 10 Assembling the Tube with the Male Connector NOTE: Tube size is recommended 3 mm-ID and 6 mm-OD and made of silicon which has thermal resistance up to at least 100ºC. 11.
  • Page 6 09931525A Close the sample compartment cover. In the spectrum FL software, click the Accessory icon to perform the calibration. The following window appears. Command Function Total Step The total accessory number of steps is fixed at 27500. Select the calibration scan speed. 30, 60, 240, 1200, 2400, 12000 Scan Speed nm/min.
  • Page 7 09931525A Set the PMT voltage and click Reset to format the Multi-Cell steps. Click Calibrate, and then the following dialog box will appear. Click OK. The Multi-Cell calibration will start. NOTE: If the peaks do not look as shown in step 8 or if any peak is saturated, click Stop, and then go back to step 5 and lower the PMT Voltage value and resume the calibration.
  • Page 8 09931525A M easurem ent 1. Run Spectrum FL software as Administrator and select a measurement mode. 2. Check the recognition of Accessory. 3. Set up the measurement parameters in turn. NOTE: For more details of method setup, refer to Spectrum FL Software Users Guide. 4.
  • Page 9: Troubleshooting

    2. If the intensity value is still low after the recalibration, the lamp of the light source may be deteriorated and needs replacing. Contact your local PerkinElmer service office. W hen the M ulti-Cell does not reset If the resetting is failed, the photo interrupter switch inside the Multi-cell Holder needs replacing.
  • Page 10 09931525A...