9.4.3 Check Weighing

During weighing of a sample the balance can be set to show if the weight is
above or below an upper and a lower limit. The display will use the arrows
under the bar graph to show that check weighing is operating.
between the arrows will indicate when the weight is below the lower limit,
between the limits or above the upper limit.
The buzzer can be set to be active when the weight is outside the limits
(below the lower or above the upper) or within the limits (above the lower and
below the upper limit), or turned off. If desired, only one limit needs to be set.
If only one limit is set the other limit is considered to be zero (lower) or the
maximum (upper).
The Check weighing is not active when the weight is less than 20d. This is
the minimum weight at which the indicator bars are displayed and the buzzer
sounds, if it is not set to OFF (see the next page).
Steps:
• Press [Mode] and then the [Up] or [Down] key to select Check
Weighing. When "CHECK" is displayed press the [Enter] key to enter
Check Weighing mode
• Press [Up] or [Down] to set the "LOW LIMIT" to "On" or "OFF"
Press [Enter] to proceed. If the "LOW LIMIT" was set to "On" the
display will show the current low limit in the unit last used in the
weighing mode. You may change the limit using the numeric entry
method (see section 7.1).
• Press [Enter] to proceed
• If the "LOW LIMIT" was set to "OFF" or the setting of the low limit is
complete, then the display will change to "HIGH LIMIT". Use [Up] and
[Down] to set the "HIGH LIMIT" to "On" or "OFF"
• Press [Enter] to proceed
• If the "HIGH LIMIT" was set to "On", the display will show the current
high limit in the last selected unit which can be changed in the same
way as in case of "LOW LIMIT"
• Press [Enter] to proceed