13.5 CALIBRATION SETUP

This menu allows the user to set the calibration parameters.
• Press [Enter] when "CAL SETUP" is displayed to select the calibration
parameters
• The options for each parameter can be scrolled through by using the
[Up] or [Down] key
ENABLE
CAL REPORT
TIME CAL
TEMP CAL
INT CAL
(Only on balances with
internal calibration masses)
• Press [Mode] to advance to setting of the next menu which is
"PASSCODES" or [Esc] to return to normal weighing.
13.5.1

INTERNAL CALIBRATION MASS ADJUSTMENT

Balance model
PGW 153i
PGW 253i
PGW 453i
PGW 753i
PGW 1502i
PGW 2502i
PGW 3502i
PGW 4502i
NOTE: The value of the internal mass is not an exact number, such as 100g.
It must be set by comparing with an external mass. The exact value can be
set, provided this feature is enabled at the factory.
© Adam Equipment Company 2007
NO =operator calibration is disabled
YES=operator calibration is enabled
On = Enabled. Prints out Calibration
report after successful calibration
OFF = Disabled
On = Enabled. Select time from 1 to 24
hours. Default setting is 4 hours
OFF = Disabled
On = Enabled. If set to On, set the
temperature variation from 0.2 to 4°C
OFF = Disabled
YES = Use Internal calibration mass as
displayed or adjust for accurate Internal
Calibration (see section 13.5.1)
NO = Use external calibration mass
Internal Calibration mass (approx.)
100g
240g
240g
520g
1400g
1400g
1400g
1400g
Page 35 of 48