"OFF" or "On". The user can enable or disable the parts counting mode
by using [Up] or [Down]. Pressing [Enter] will confirm the setting and
will advance to the next weighing mode. Repeat for each mode in turn
• Press [Mode] to advance to setting of the next menu or press [Esc] to
return to normal weighing

13.3 ENABLE SERIAL INTERFACE PARAMETERS

The parameters affecting the serial setup are set in a similar manner to the
other parameters.
Press [Enter] when "SERIAL" is displayed to enter the sub-menu.
The parameters that can be set are:
ENABLE
BAUD RATE
PARITY
STABLE
CONTINUOUS Set the RS-232 to send data continuously to On or OFF
PERIODIC
FORMAT
Format of custom forms #1 and #2
If FORM1 or FORM2 is selected, it can be changed by the user using a
selection of available data. By default the 2 forms are the same as the
standard form unless changed by the user as below.
© Adam Equipment Company 2007
The serial port can be set to On or OFF
Set the Baud Rate to 4800, 9600, 19200 or 38400, the
default rate being 4800
Set the Parity to NONE, EVEN or ODD
To print when stable (On) or regardless of stability (OFF)
Set the RS-232 to send data periodically (set in seconds)
to On or OFF. If On is selected, the value can be changed
between 1 and 999 seconds, using [Up] and [Down]
To send data as a single line of data, using the standard
format or using a customer designed format (FORM 1 or
FORM 2).