6 Save the calculation as described in
To calculate LV Mass
1 On a frozen 2D image, press the CALCS key.
2 In the calculations menu, locate LV mass.
3 Do the following for EPI and then for Endo:
a Select the measurement name from the calculations menu.
b Position the caliper where you want to begin the trace, and press the SELECT key.
c Using the touchpad, trace the desired area.
i To make a correction, select Undo on-screen or press the backspace key.
d Complete the trace, and press the SET key.
e Save the calculation.as described in
4 Select Apical from the calculations menu.
5 Positioning the calipers, measure the ventricular length. See
6 Save the calculation.
To measure peak velocity
For each cardiac measurement, the system saves up to five individual measurements and calculates their
average. If you take more than five measurements, the most recent measurement replaces the fifth one. If you
delete a saved measurement from the patient report, the next measurement taken replaces the deleted one
in the patient report. The most recently saved measurement appears at the bottom of the calculations menu.
1 On a frozen Doppler spectral trace, press the CALCS key.
2 From the calculations menu, select MV, TV, TDI, or P. Vein.
3 Do the following for each measurement you want to take:
a Select the measurement name from the calculations menu.
b Position the calipers.
c Save the calculation as described in
88
"To save a calculation"
on page 74.
"To save a calculation"
"To save a calculation"
on page 74.
"Working with calipers"
on page 74.
on page 67.