Table 9-4: Manufacturer's Declaration - Electromagnetic Immunity (continued)
Immunity Test
IEC 60601 Test Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
(continued)
Note
U
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the FUJIFILM SonoSite ultrasound
system is used exceeds the applicable RF compliance level above, the FUJIFILM SonoSite ultrasound system should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the FUJIFILM SonoSite ultrasound system.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
164
Compliance Level
3 Vrms
is the AC mains voltage prior to application of the test level.
T
Electromagnetic Environment
Portable and mobile RF
communications equipment
should be used no closer to any
part of the FUJIFILM SonoSite
ultrasound system including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended Separation
Distance
= 1.2 √
d
P
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic Site survey
should be less than the
compliance level in each
b
frequency range
.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
(IEC 60417 No. 417-IEC-5140:
"Source of non-ionizing radiation")
Electromagnetic compatibility
a
,