WARNING
158
Unless following ESD precautionary procedures, do not connect to or touch (with
body or hand-held tools) pins (contacts) of connectors that have the ESD
Sensitive Devices label:
Electrostatic Sensitive
Devices Label
Pins (contacts) on transducer
and back of TTC connector
Pins (contacts)
on the system
If the symbol is on a border surrounding multiple connectors, the symbol pertains
to all connectors within the border.
ESD precautionary procedures include the following:
All staff involved must receive training about ESD, including the following at a
minimum: an explanation of the ESD warning symbol, ESD precautionary
procedures, an introduction to the physics of electrostatic charge, the voltage
levels that can occur in normal practice, and the damage that can occur to
electronic components if equipment is touched by an individual who is
electrostatically charged (IEC 60601-1-2, section 5.2.1.2 d).
Prevent the buildup of electrostatic charge. For example, use humidification,
conductive floor coverings, nonsynthetic clothing, ionizers, and minimizing
insulating materials.
Discharge your body to earth.
Use a wrist strap to bond yourself to the ultrasound system or to earth.
Pins (contacts) on the back of the
Triple Transducer Connect (TTC)
Electromagnetic compatibility