Table 10-4: Transducer Surface Temperature Rise, External Use (°C)
Test
C11x
rC60xi
Still air
14.2
15.0
Simulated
7.3
8.9
Use
Table 10-5: Transducer Surface Temperature Rise, Internal Use (°C)
Test
ICTx
Still air
9.2
Simulated
5.2
Use

Acoustic output measurement

Since the initial use of diagnostic ultrasound, the possible human biological effects (bioeffects) from
ultrasound exposure have been studied by various scientific and medical institutions. In October 1987, the
American Institute of Ultrasound in Medicine (AIUM) ratified a report from its Bioeffects Committee (Bioeffects
Considerations for the Safety of Diagnostic Ultrasound, J Ultrasound Med., Sept. 1988: Vol. 7, No. 9
Supplement). The report, sometimes referred to as the Stowe Report, reviewed available data on possible
effects of ultrasound exposure. Another report, "Bioeffects and Safety of Diagnostic Ultrasound, " dated
January 28, 1993, provides more current information.
The acoustic output for this ultrasound system has been measured and calculated in accordance with
"Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment" (NEMA UD2-2004), and
"Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound
Equipment" (NEMA UDe3-2004).

In Situ, derated, and water value intensities

All intensity parameters are measured in water. Since water does not absorb acoustic energy, these water
measurements represent a worst case value. Biological tissue does absorb acoustic energy. The true value of
the intensity at any point depends on the amount, type of tissue, and the frequency of the ultrasound passing
through the tissue. The intensity value in the tissue, In Situ, has been estimated by using the following formula:
-(0.23alf )
In Situ= Water [e
Acoustic output measurement
HFL38xi
HFL50x
12.4
10.7
7.7
7.7
TEExi
8.6
4.7
]
HSL25x
L25x
L38xi
17.5
16.1
12.5
9.1
8.5
8.8
P10x
rP19x
16.0
14.9
9.1
7.6