3 Do the following for each measurement you want to take:
a From the calculations menu, select the measurement name.
b For twins, select Twin A or Twin B, and then select the measurement name.
Note
c Position the calipers.
d Save the calculation as described in
To measure fetal heart rate (M Mode)
1 On a frozen M Mode trace, press the CALCS key.
2 Select FHR from the calculations menu. A vertical caliper appears.
3 Using the touchpad, position the vertical caliper at the peak of the heartbeat.
4 Press the SELECT key. A second vertical caliper appears.
5 Using the touchpad, position the second vertical caliper at the peak of the next heartbeat.
6 Save the calculation as described in
Menu Heading
MCA (Middle Cerebral Artery)
Umb A (Umbilical Artery)
*Calculation requires a trace measurement.
To calculate MCA or Umba (Doppler)
Note
1 Select OB exam type, and select LMP or Estab.DD in the patient information form.
2 On a frozen Doppler spectral trace, press the CALCS key.
3 Do the following for each measurement you need to take:
Exam-based calculations
The caliper tool may change depending on the measurement selected, but the
position remains constant.
"To save a calculation"
"To save a calculation"
The system does not provide an MCA/UmbA ratio from the PI (Pulsatility Index).
on page 74.
on page 74.
OB Calculation
S/D, RI
S/D, RI, PI*
S/D, RI
S/D, RI, PI*
Results
SD, RI
SD, RI, PI
SD, RI
SD, RI, PI
99