Calibration Of Frequency Meter - ACP Speecon 7200M3 Manual

Table of Contents

Advertisement

(1mA) full-scale, 3V full-scale output is used at
maximum output frequency [constant (Pn-02)]
operation.Set constant (Pn-45)=0.30.
Description
Frequency reference value is displayed.
DSPL
DATA
PRGM
STR
Select PRGM mode.
Pn-01
DRIVE
RESET
Select constant (Pn-45 )
Pn-45
DATA
STR
Display Pn-45 data.
1.00
Change the set value
0.30
Write in the set value. ("End is displayed)
The data are displayed
0.30
Note:
1. Since analog monitor gain is set to 1.00 prior shipping, 10V is output at maximum output frequency
[constant (Pn-02)] operation.
2. By data display of constant (Pn-45) in the program mode, voltage at 100% level according to the
constant (Pn-45) set value is output by the meter calibrating function without any conditions.
(Example) Assuming constant(Pn-45)=0.30 : 10V×0.30=3V is output without any conditions.
Key Operation
Digital Operator Display
- -
59
Remarks
LED DRIVE OFF
(Note)
Displayed for 0.5 second.
Confirm the display.